MAN205-0195042EN Rev. A Aug. 2011
Page 33 of 137 pages
Measuring steam generator pressure
The measurement range is 100-600kPa (14.5–87 psi)
The resolution measurement and display is 1kPa (1 psi).
The accuracy of the measurement is a maximum 1% of the range, i.e. 5kPa (0.7 psi)
Calibration of the pressure circuit requires a special engineer access code.
3 Serial Ports
The control system contains 3 serial ports (1optional) according to the following
specifications:
2 Serial Ports for the connection of up to 2 command panels.
1 Serial Port connected to an external PC computer (RS232/485).
1 Serial Port (optional)
Serial Port RS232 is included in the system to connect to an external PC up to a
distance of 8 meters. If it becomes necessary to convert certain functions or to
reprogram or reload software of the autoclave from a distance of over 8 meters, a
modem and Serial Port RS485 is required. This enables service technicians to attend to
the unit from a remote site. ADMC software enables full control of the autoclave.