Running a Cycle
Before starting a cycle, make sure to check the following:
Select the steam source.
If the building steam source is supplying enough steam to operate the
autoclave. If not, use the internal steam generator. See ―Switching to
the Steam Generator‖ section in the ―Operating Instruction section.
There is enough steam pressure in the jacket.
There is enough air pressure for the pneumatic valves.
The door is closed.
There is enough distilled or mineral free water for the steam generator.
There is enough water in the vacuum pump water reservoir.
When the SETUP has been checked and the display is in the Stand By /Ready mode,
press START to run the cycle; the RUN LED lights up and the cycle is in progress.
At the end of the cycle, the LED switches OFF, and the appliance is in the Stand By
mode.
Powering ON the Reset Test
1. Turn ON the appliance.
2. Check that all the valves and outputs, including LEDs on the panel are turned
OFF.
3. Check the last illumination of the panels is turned ON.
4. Check the message Wait!!! Version = (version name) appears on the 2
panels with the correct version.
5. Select two-digit program numbers by pressing each digit-key within 2 seconds
of each other.
6. Check that the two LEDs on each panel are turned OFF, the message deleted
and the system moves to into the Stand By Mode.