MAN205-0195042EN Rev. A Aug. 2011
Page 60 of 137 pages
Legend
V - Air removal stage
S - Sterilization stage
F - Fail
E - Exhaust stage
H - Heating stage
D Drying stage
PRINTER OUTPUT
DESCRIPTION
Autoclave 1102058
Autoclave serial no.
Name : __--------
Name of the operator to be filled in .
Load #: 0030
30/08/2011 13:45:16
Date and time the cycle ended
E13: 31 099.7ºC 049.5k
The time, temperature and pressure during
exhaust stage.
F11:30 134.2ºC 306.8k
If cycle fails, conditions at time of failure.
CYCLE FAILED! ! !
Message displayed if cycle fails.
Manual Stop - 03
If operator presses the manual stop button.
H11:28 099.9ºC 26k
The time, temperature and pressure during
sterilization.
H10:08 134.3ºC 315.1k
The time, temperature and pressure during the
heating stage.
H06:04 132.3ºC 304.1k
The time, temperature and pressure during the
heating stage.
H02:00 116.2ºC 217.3k
The time, temperature and pressure during the
heating stage.
H03:00 080.6ºC 023.1k
The time, temperature and pressure during the
heating stage.
V00:27 078.3ºC 048.7k
The time, temperature and pressure during air
removal (vacuum).
Exhaust mode: 1
Exhaust mode: fast exhaust and process ends..
Dry time : 001 min
Drying time of the selected program
Ster time: 001.0 min.
Preset sterilization time of the selected program.
Ster Temp: 134ºC
Sterilization temperature in chamber for selected
program.
Cycle: 01 Unwrap
Selected program: Unwrapped Instruments.
30/08/2011 13:29:31
Date/month/year time/min./sec.
Version = Horiz En3M3 Spn
Number and version of the software program.

Table of Contents