Operating Analog Output
The analog outputs are set in the factory according to the order form.
If you want to change the settings, follow these instructions. See
section Analog Output Module on page 38.
Changing Output Mode and Range
The output channel has its own DIP switch module with 8 switches,
see the position in Figure 22 on page 38 (DIP switches to select the
output mode and range). To change output mode and range:
1. Select
the current/voltage output, switch ON either of the
switches, 1 or 2.
2. Select the range, switch ON one of the switches from 3 to 7.
Figure 41 Current/Voltage Switches of Output Module
The number refers to Figure 41 above:
1 = Current/voltage selection output switches (from 1 to 2)
2 = Current/voltage range selection switches (from 3 to 7) in
analog output 1 and 2.
3 = Switch for service use only. Keep in OFF position always.
NOTE
Only one of the switches 1 or 2, must be ON at a time. Only one of
the switches 3 to 7, must be ON at a time.