Chapter 3 _______________________________________________________________ Installation
VAISALA_______________________________________________________________________ 39
9. If you are replacing an existing module, the installation is
complete. If you are adding a new module, continue to the next
step.
10. Connect to the service port, and execute the MODS command to
register the changes.
11. Switch the power off and back on again to clear related errors
and to activate the changes.
Figure 23 Analog Output Dip Switch Positions
NOTE
For instructions on how to operate the analog output, see section
Operating Analog Output on page 102.
NOTE
Only one of the switches 1 and 2 can be ON at a time. Only one of
the switches 3...7 can be ON at a time.