Table 27 Mechanics
Cable bushing M20x1.5 For cable diameter 8...11mm/0.31..0.43"
Conduit fitting 1/2”NPT
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification IP 65 (NEMA 4)
Option1 with plug (female) with 5m/16.4ft
black cable
User cable connector
(optional)
M12 series 8- pin (male)
Option2 with plug (female) with screw
terminals
D9-connector
Technical Specifications of
Optional Modules
Table 28 AC Power Supply Module
Operating voltage 100...240 VAC 50/60 Hz
Connections Screw terminals for 0.5...2.5 mm2 wire (AWG 20...14)
Bushing For 8...11 mm diameter cable
Operating temperature -40...+60°C (-40...+140°F)
Storage temperature -40...+70°C (-40...+158°F)
Table 29 Analog Output Module AOUT-1
Outputs 0...20 mA, 4...20 mA, 0...1 V, 0...5 V, 0...10 V
Operating temperature range -40...+60 °C (-40...+140 °F)
Uout 0...1 V max 30 mA
Uout 0...5V/0...10V max 30 mA
Power consumption
Iout 0... 20 mA max 60 mA
External loads current outputs RL< 500 ohms
540 ohms
0....1 V RL> 2000 ohms
Max load + cable loop resistance
0...5 V and 0... 10 V RL> 10 000 ohms
Storage temperature range -55...+80 °C (-67...+176 °F)
3-pole screw terminal
Max wire size 1.5 mm
2
(AWG16)