Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 79
Example of setting the limit to 0.5 hPa:
>dpmax <cr>
Pd max : 1.000 ? 0.5 <cr>
>
Use the ERRS command to analyze problems
AVRG
Use the AVRG command to set barometer measurement averaging
time (in seconds). Valid range of averaging time is 1...600 seconds or
0, which disables the measurement averaging function.
AVRG [x] [?]
where:
x = Average time
Example:
>avrg
Average filter : 1 s ?
>
HHCP
The HHCP command shows or sets the altitude for height corrected
pressure. The valid range of HCP height is -30... 30 m.
HHCP [?] [x] [y]
where:
? = Displays the current altitude offset for HCP calculation. You
cannot use this modifier with the other modifiers.
x = Numeric value (altitude).
Ranges:
-30 … 30 m
-99… 99 ft.
y = Distance scale used.
Values:
m
ft