Chapter 3 _______________________________________________________________ Installation  
VAISALA_______________________________________________________________________ 33 
Installation 
To install the AC power supply module: 
1.  Disconnect the power and open the barometer cover. 
2.  Remove the protective plug from the cable gland and thread the 
wires. If the power supply module is installed in the factory, 
continue with the step 5. 
3.  To attach the module fasten the power module to the bottom of 
the housing with four screws. See the position of the power 
supply module in Figure 15 on page 25. 
4.  Connect the wires from t
he terminals of the power supply 
module marked with + and - to the terminals POWER 10 ... 35 
V on the motherboard of the barometer. 
5.  Connect the AC mains voltage wires to the power supply 
module terminals marked with N and L. 
6.  Attach the grounding wire to the grounding terminal on the 
right-hand side of the barometer.   
7.  Connect the power. The LED on the cover of the barometer is lit 
continuously during normal operation. 
 
WARNING 
Do not detach the power supply module from the barometer when the 
power is on. 
 
WARNING 
Do not connect the mains power to power supply module when it is 
not installed in the barometer. 
 
WARNING 
Always connect protective ground terminal.