Chapter 6 ___________________________________________________Calibration and Adjustment  
VAISALA______________________________________________________________________ 123 
Feeding Adjustment Information 
This information is shown on the device information fields, see 
section Information Display on page 53 and the ? command on page 
87. 
Using Display/Keypad 
To feed in the adjustment information: 
1.  If you are not in the adjustment menu, press the 
ADJ button on 
the motherboard to open the 
ADJUSTMENT MENU. 
2. Select 
Adjustment info, press the right arrow button. 
3. Select 
Date, press the SET button. Enter date by using the arrow 
buttons. Press the 
OK button. 
4. Select 
i, press the SET button. Enter information text including 
17 characters at maximum. Use the arrow buttons. Press the 
OK 
button. 
5. Press the 
EXIT button to return to the basic display. 
Using Serial Line 
CTEXT 
Use the CTEXT command to enter text to the adjustment information 
field.  
Example: 
>ctext Vaisala/MSL 
Calibration text : Vaisala/MSL 
> 
 
CDATE 
Use the CDATE command to enter date to adjustment information 
field. Set the adjustment date in format YYYY-MM-DD. 
>cdate 2006-06-12 
Calibration date : 2006-06-12 
>