PA11A Message Format 
Message type 1 format is as follows: 
<sp>P1<sp>P2<sp>P3<sp>status<sp>average<sp>trend<cr><lf> 
where:   
<sp> = Space character 
P1  =  Pressure output from transducer 1 (in 0.1 hPa). The 
reading is written with 5 characters. 
If the transducer has a fault or has been switched off, the 
reading is replaced with error status /////. 
P2  =  Pressure output from transducer 2, same format as P1 
P3  =  Pressure output from transducer 3, same format as P1 
status  =  Status of the average pressure measurement (binary, 
with 8 characters). 10000000 means OK with all three 
transducers included. Otherwise, the last three bits of the 
status indicate the transducers that are included when 
calculating the average. For example, status 00000110 
means that transducers 2 and 3 are included and 
transducer 1 is ignored. 
average  =  Average pressure (with 5 characters) based on three 
consequent measurements of each transducer included, 
i.e. pressure average over approx. one minute. 
If the average pressure cannot be calculated, the reading 
is replaced with error status /////. 
trend  =  Three hour pressure trend (with 3 characters). Falling 
trend is indicated with a minus "-" character before the 
reading. If the data is not available (when measurement 
has been running for less than three hours), the reading 
is replaced with ///. 
<cr><lf>  =  Carriage return and line feed characters 
 
NOTE 
All pressure readings are in 0.1 hPa. The pressure unit selection has 
no effect. 
If a reading is shorter than the amount of characters reserved for it, it 
is left-padded with space characters. See the examples below.