USER'S GUIDE____________________________________________________________________  
116 _________________________________________________________________ M210855EN-D 
Linear/Offset Adjustment Using 
Serial Line 
NOTE 
Making adjustments is possible only after adjustments have been 
unlocked. To unlock the adjustments, press the ADJ button on the 
motherboard of the transmitter. 
 
LCP1, LCP2, LCP3 
The LCP1 command performs offset adjustment for barometer 
module/module P1. The 
LCP2 and LCP3 commands do the same for 
modules P2 and P3. Use the commands: 
-  to view current linear/offset adjustments 
-  to activate or deactivate the linear/offset adjustment function 
-  to enter new offset or linear offset/gain pressure adjustments to the 
transmitter. 
NOTE 
Each barometer module has its own adjustment commands. You must 
enter the linear/offset adjustments for each barometer module 
separately. 
 
To enter new offset (1-point) or linear (2-point) adjustments: 
1.  Enter the command 
LCP1 ? (and LCP2 ?, LCP3 ?, if 
applicable) to display current pressure adjustments. 
2.  Write down the current settings if you want to restore the old 
values later on. 
3.  Enter the command 
LCP1 OFF (and LCP2 OFF, LCP3 OFF, 
if applicable) to deactivate the previous adjustments. 
4.  Put PTB330 to reference pressure conditions and write down the 
reference and measured P
1
 (P
2
, P
3
) values for linear offset/gain 
adjustment.