Chapter 6 ___________________________________________________Calibration and Adjustment  
VAISALA______________________________________________________________________ 115 
After disabling the previous adjustments, calibrate the PTB330 against 
your pressure reference. Make a note of the pressure reading of each 
PTB330 barometer module separately (quantities P
1
, P
2
, and P
3
). Then 
use the display/keypad to enter and activate the new multipoint/linear 
adjustment factors: 
1. Press the 
ADJ button located on the motherboard in the upper 
corner on the left-hand side to open the 
ADJUST PTB330 
menu. 
2. Select 
P1 adjustments, press the right arrow button. 
3. Select 
Multipoint/Linear adjustment, press the SET button. 
4.  Select an existing adjustment point or
 Add adjustment point 
and press the 
SET button to enter your new values. If asked, 
select 
MODIFY to replace the old values with your new ones. 
5.  Enter the P
1
 value of the PTB330 at the Reading prompt using 
the up/down arrow buttons. Press the
 OK button. 
6.  Enter the corresponding pressure reference value to the 
Reference prompt using the up/down arrow buttons. Press the 
OK button. 
7.  Repeat steps 4…6 until you have entered all the pressure points 
you have. 
8.  To remove any excess old adjustment points, press 
SET and 
select 
REMOVE on each of them. 
9.  Once you have finished, press the 
OK button. Press the YES 
button to activate the adjustment you have configured. 
10.  Press the left arrow button to go back to the 
ADJUST PTB330 
menu and repeat steps 2…9 for each barometer module you are 
adjusting. 
11.  If necessary, go to the Adjustment Information submenu and set 
a new calibration date and note. 
12.  Exit the adjustment mode by pressing 
EXIT.