Chapter 6 ___________________________________________________Calibration and Adjustment  
VAISALA______________________________________________________________________ 117 
5.  For linear adjustment, enter the command LCP1 (LCP2, 
LCP3), and enter the new adjustments when prompted. See the 
first example. 
For 1-point offset adjustment, enter the command 
LCP1 (LCP2, 
LCP3), and enter the measured and reference values in the first 
slots, and then enter the same values increased by one in the 
second slots. See the second example. 
6.  Enter the command 
LCP1 ON (and LCP2 ON, LCP3 ON, if 
applicable) to activate the new adjustments. 
LCP1 [x/y] [z] 
where: 
x = ON 
y = OFF 
z = ? 
 
Examples: 
Linear adjustment: 
>lcp1 
1. Reading ? 981.2 
1. Reference ? 980.0 
2. Reading ? 1102.1 
2. Reference ? 1100.0 
> 
 
Offset adjustment: 
 
>lcp1 
1. Reading ? 981.2 
1. Reference ? 980.0 
2. Reading ? 982.2 
1. Reference ? 981.0 
> 
 
NOTE 
The new linear/offset adjustments cancel the previous linear/offset 
adjustments as well as the valid date of calibration of the transmitter. 
A new calibration date and note can be set with the CDATE and 
CTEXT commands.