Figure 74 QMR102 Adjustment and the Foam Location
3. It is important to ensure that the rim of the rain gauge is leveled
precisely. Failure to do this will result in a systematic error. Use
a spirit level (1) and adjust with the fixing screws (3). See
Figure 74 above.
4. The cable length can be shortened or lengthened as required. If
the cable is lengthened, please ensure a good quality
environmental connector, or a heatshrink joint (see Figure 75
below). Extension cables used must be of a similar specification.
9902-004
Figure 75 Wiring Diagram of QMR102
NOTE
When using QMR102, the shield must be connected to the ground.