What to do if VAT 612 Control Unit pressure is not 0 after ZERO?
- GGregory BaileyAug 26, 2025
If the pressure reading is not "0" after performing ZERO on your VAT Control Unit, wait for the sensor voltage to stabilize before performing ZERO. Ensure the system is pumped to base pressure and the valve is open before performing ZERO. If the sensor offset voltage exceeds ±1.4V, replace the pressure gauge.