10. Technical Specifications
Green X / X18
(Model: PHT-75CHS) Installation Manual 133
English
10.3 Electrical Specifications
Item Description
Power supply voltage 100 - 240 V ~
Frequency 50 / 60 Hz
Power rating 2.2 kVA
Accuracy
Tube Voltage (kVp) ± 10 %,
Tube Current (mA) ± 20 %
Exposure Time (s) ± (5 % + 50 ms)
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ±10 %.
Mode of operation: non-continuous operation (NFPA 70: long time
operation) - needs waiting time (at least 60 times the exposure time) before
the next exposure.
Column operation time: Max. 2 min. On / 18 min. Off (Ratio 1:9)