26 DSGH Installation and Operation Guide
Installation
You can calibrate the current loop output on the bench before mounting
the detector on the process. See page 3-43.
GEN2000 terminals 13 and 14
250 – 800 
load resistor
(optional)
Mini
clips
HART
modem
RS-232
cable
PC running
VEGA software
Transmitter
test points
H1
H2
Bench test setup
Note: You may need to reset the time and date if the gauge has not
had power for > 28 days. The Real Time Clock Fail message
may appear. You must enter the correct time and date. The
clock is the basis for source decay calculations.