52 DSGH Installation and Operation Guide
Calibration
Standardization on process
This method requires measuring the process in the pipe with the density
gauge and entering the laboratory measured density into the software.
Advantages: Standardization on process is the easiest method. It ensures
the density gauge reads what the laboratory reads at that density.
Disadvantages: It is not the most accurate or repeatable method. In
addition, it requires a laboratory for results.
Standardization on absorber plates
Absorber plates have the following features:
Usually made of lead.
Insert into slots in front of the gauge.
Require a mounting kit from VEGA.
Require the use of the same plates to ensure a consistent absorption of
radiation (this is a substitute for the radiation absorption of the material in
the pipe.
Requires emptying process from the pipe and access to the measuring
assembly
Note: Select Lab Sample on the Setup | Cal Parameters tab.
Note: This method requires you to select Default Value and enter the
Standardize Default Value on the SetUp | Cal Parameters tab.