DSGH Installation and Operation Guide 49
Calibration
2 Setting the high density and collecting Cal High data
You must:
Use the gauge to measure the high process condition.
Enter the actual density.
This sets the gain of the calibration curve. Perform this procedure before or
after setting the low density.
Before starting the Cal High data collection:
Check that the correct parameters (vessel inner diameter, engineering
units, measurement span, and source type) are correct.
Power up gauge one hour before start of calibration.
Fill vessel or pipe with high process or close the source holder shutter to
simulate high process.
Prepare to draw a sample while the gauge is collecting data.
Procedure 3.5: To set the cal high density
3. Click Accept.
4. Enter the actual value in engineering units.
5. Click OK.
1. Select Calibration | 2 Point Calibration | Cal High Collect.
2. Select Start.
3. Click Accept.
4. Enter the actual density process value (from the laboratory)
in engineering units.
5. Click OK.