DSGH Installation and Operation Guide 87
Diagnostics and repair
Source Functions
Recording the source wipe and shutter check
You can use the gauge’s diagnostic alarms to remind you when a source
wipe and shutter check are due. If you do, you must record the source
wipes and shutter checks in the software to acknowledge the alarm and
reset the timer.
Perform this procedure after a source wipe or a shutter check.
Procedure 5.1: To record a source wipe or shutter check
Note: See the Radiation Safety for U.S. General and Specific
Licensees, Canadian and International Users Manual and the
Radiation Safety Manual Addendum of Reference Information
CD that came with the source holder and the appropriate
current regulations for details.
1. Select Setup | Source Functions.
2. Click Record Wipe or Record Shutter Check.