DSGH Installation and Operation Guide 65
Advanced functions
Auxiliary Input Test
This mode simulates the auxiliary input frequency at a user-defined number
of counts. The effect of auxiliary input counts depends on the auxiliary input
mode.
Examples:
• Temperature probe
•Flow meter
• Second transmitter
While in this mode, after entering a number of counts, it may be useful to
look at the Process Chain tab to view the variables affected by the auxiliary
input counts value.
Procedure 4.7: To perform an auxiliary input test
3. Remove the gauge from control. Enter the value of the new
counts to force.
4. Click OK.
The transmitter functions in this mode until it times out (1
hour), or you click Exit and OK.
1. Select Diagnostics | Test | Auxiliary Input Test.
2. Click Enter.
3. Remove the gauge from control. Enter the auxiliary counts.
4. Click OK.
The transmitter functions in this mode until it times out (1
hour), or you click Exit and OK.