64 DSGH Installation and Operation Guide
Advanced functions
Procedure 4.5: To perform a current loop test
Sensor Test
This mode simulates the sensor output at a number of raw counts you
define. This is before application of:
Temperature compensation
Sensor uniformity gain
Standardize gain
The true sensor output is ignored while the transmitter is in sensor test
mode.
This mode is useful for verifying the electronics and software response to
input counts without having to:
Change the process
Shield the source
Vary the radiation field
While in this mode, after entering a number of counts, it may be useful to
look at the Process Chain tab to view the variables affected by the raw
counts value.
Procedure 4.6: To perform a sensor test
1. Select Diagnostics | Test | Current Loop Test.
2. Click Enter.
3. Remove the gauge from control.
4. Enter the current loop test value.
5. Click OK.
The transmitter functions in this mode until it times out (1 hour), or you
click Exit and OK.
1. Select Diagnostics | Test | Sensor Test.
2. Click Enter.