DSGH Installation and Operation Guide 59
Advanced functions
Gauge Info tab
Table 4.3
Gauge Info tab — additional display values
Uniformity Gain Displays the current value of the uniformity gain. Use this
to force all density gauge sensors to output the same
counts at a given radiation field
Source Decay Gain The current value of the source decay gain. Use this to
compensate for the natural decay of the radiation source,
which produces a lower field over time.
Stdz Gain The current value of the standardize gain that adjusts with
each standardize procedure.
HV Setting The set point for the sensor high voltage.
Table 4.2
Process Variables tab — display values
Value Description
Scintillator sensor voltage
Firmware’s version on the
FLASH
Hardware’s version number
GEN2000 CPU board’s
serial number
GEN2000 unit’s serial
number
Sensor Coefficients
T0 – T3
Value Description
Sensor
Coefficients
The algorithm that compensates for variations in measurement
output with changes in temperature uses temperature
coefficients. The factory determines the coefficients through
rigorous testing. You cannot change them through normal
operation.