Sterilizer tests
48 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
CARRY OUT THE TEST
1 Empty the sterilizer chamber to ensure no load is present.
2 Close the chamber door and ensure the sterilizer chamber is completely dry and
cold to avoid any false negative outcome.
3
On the homepage, tap Vacuum test.
4 Tap and enter your credentials if required: the chamber door locks.
5
Wait until the end of the test and tap OPEN: the chamber door unlocks.
6 Enter your credentials if required:a message informs if the test passed or failed. If
the test failed, see "What to do when the test failed" below
WHAT TO DO WHEN THE TEST FAILED
1 Check, clean or replace the door gasket.
2 Clean the chamber face side and the chamber filters.
3 Repeat the Vacuum test. See "Carry out the test" above.
4 If the test fails repeatedly, call technical service.