Cleaning the ion block, isolation valve, and extraction cone 5-35
11. Holding the PEEK terminal block gently, use the needle-nose pliers to
gently grasp the heat-shrink tubing on the heater cartridge assembly
and slide it and the PEEK terminal block out of the ion block.
12. Use the O-ring removal kit to carefully remove the cover seal from the
ion block (see also “Removing O-rings and seals” on page 5-9).
TP02618
Heat-shrink tubing
Heater cartridge assembly
PEEK terminal block
TP02619
Cover seal