Set up unit 17
EN
Bacteria filter
If the VENTIlogic is intended for use by several patients
(e.g. in hospital), bacteria filter WM 24148 should
be used to protect against infection. It is installed be-
tween the hose system and the VENTIlogic or the
VENTIclick. Please also observe the enclosed operat-
ing instructions for the bacteria filter.
Caution!
The bacteria filter represents an additional resistance in the air stream. This may
lead to a change in the response characteristics of the trigger. For this reason,
you should have the unit parameters reset if a bacteria filter is fitted retrospectively.
VENTIpower
For mains-power independant operation
VENTIpower
is available.
VENTIpower can also be connected to
VENTIlogic
in
parallel to mains power. VENTIpower then remains in
standby mode. If mains power fails, VENTIpower au-
tomatically supplies power to the therapy device with
a delay of about 4 seconds.
Using VENTIpower you can operate VENTIlogic inde-
pendant from mains power in the carrying bag
WM 24888. To do so place VENTIpower in the ded-
icated section of the carrying bag and connect it to the
therapy device.
Please also observe the operating instructions for
VENTIpower.