46 Technical specifications
EN
11. Technical specifications
VENTIlogic VENTIlogic with VENTIclick
Product class as per EC Directive
93/42/EEC
IIa
Dimensions WxHxD in cm 23 x 12.5 x 34 23 x 12.5 x 45.5
Weight approx. 4.5 kg approx. 4.8 kg
Temperature range
operation
storage
+5 ˚C to +35 ˚C
–40 ˚C to +70 ˚C
Air pressure range 600 – 1100 hPa
Electrical connection
115 – 230 V AC, 50–60 Hz
Tolerance -20 % +10 %
Power consumption during
operation
standby
230 V 115 V
0.2 A 0.4 A
0.043 A 0.068 A
230 V 115 V
0.35 A 0.7 A
0.043 A 0.068 A
Classification as per
EN 60601-1
type of protection from electric
shock
level of protection from electric
shock
Protection class II
Type B
Electromagnetic
compatibility (EMC) as per
EN 60601-1-2
radio interference suppression
immunity to interference
EN 55011
EN 61000-3-2,EN 61000-3-3, EN 61000-4-2 to 6,
EN 61000-4-8, EN 61000-4-11
Mean sound pressure level/
operation as per
EN ISO 17510
at a distance of 1 m from unit in
patient position
approx. 35 dB (A) at 20-35 hPa
approx. 31 dB (A) at 20 hPa
approx. 29 dB (A) at 15 hPa
approx. 27 dB (A) at 12 hPa
approx. 25 dB (A) at 10 hPa
approx. 23 dB (A) at 7 hPa
Sound pressure level, alarm approx. 62 dB (A)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals