EasyManua.ls Logo

Weinmann VENTIlogic BiLevel ST - 6 Hygienic Preparation; Intervals

Weinmann VENTIlogic BiLevel ST
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 Hygienic preparation
EN
6. Hygienic preparation
6.1 Intervals
You should check the filters and wipe clean the housing and the filter compartment
cover regularly. The headgear/headband can be washed as necessary. Please also
observe the following intervals.
Interval Activity
Daily
Clean mask
Clean hose system
Clean bacteria filter in accordance with the operating
instructions
Clean VENTIclick humidifier in accordance with the relevant
operating instructions.
Clean the exhalation system after every use in accordance with
the operating instructions .
Every 24
operating hours
Change the particle filter in the bacteria filter
Weekly
Clean the mask thoroughly in accordance with the operating
instructions.
Clean coarse dust filter
Every 1000
operating hours
Change fine filter (filter change indicator ). If it is soiled,
change it sooner
Every 6 months
Change coarse dust filter. Change it sooner if soiled or worn
Change pressure measuring hose – sooner if soiled – (see “9.3
Change pressure measuring hose” on page 42).
Annually Change hose system.

Related product manuals