Hygienic preparation 25
EN
6.2 Cleaning
Clean hose system
1. Detach the hose system from the unit and the exhalation system.
2. Remove one end of the pressure measuring hose
(shake gently if necessary) and seal it with the seal-
ing plug provided. At the other end, use the sec-
ond sealing plug to seal the small opening of the
adapter to prevent water getting inside.
3. Clean the creased hose so no residues remain using a small amount of detergent
in warm water. Rinse the inside of the hose thoroughly in the process.
4. Rinse both the inside and the outside of the creased hose thoroughly using clean
warm water.
5. Shake the hose system thoroughly.
6. Hang up the hose system and allow to drip to prevent moisture getting into the
VENTIlogic.
7. Remove the plugs from the pressure measuring hose.
Dry the hose system using the VENTIlogic
1. If water accidentally gets into the pressure measur-
ing hose, insert the red drying adapter supplied in
the unit outlet.
2. Insert the adapter of the hose system in the unit out-
let or if necessary, onto the red drying adapter.
3. To activate the drying process, press the on / off
switch . Use the menu switch to reach the
patient menu. Press the dial to start the drying
process. The unit now dries the hose system.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals