7. Function check
7.1 Intervals
Perform a function check at least every 6 months. If you discover defects during the
function check, you must not use the VENTIlogic again until these defects have been
rectified.
7.2 Method
1. Assemble the VENTIlogic so that it is ready to function with the hose system, ex-
halation system and the supply cable connection.
2. Close the opening of the hose system, for example using your thumb or hand.
3. Switch on the unit by pressing the on / off switch .
4. If TA mode is active, wait approximately 4 minutes until the analysis phase is
complete. Then perform the checks described below.
5. If soft start is switched on, switch it off by pressing the soft start switch .
Depending on the operating mode set, now check the following functions:
If the values/functions set out below are not met/fulfilled, send the unit to your dealer
or to the manufacturer (WEINMANN) for repair.
Function: / Mode: CPAP S T ST SX SXX TA
Pressure setting
Respiratory frequency/
Minimum frequency
Triggering
Soft start
Flow sensor
Alarms

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals