EasyManua.ls Logo

Weinmann VENTIlogic BiLevel ST - 9 Maintenance; Intervals

Weinmann VENTIlogic BiLevel ST
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance 39
EN
9. Maintenance
9.1 Intervals
We recommend only having servicing and maintenance performed by the
manufacturer (WEINMANN) or authorized dealers.
Both filters must be regularly checked for soiling
The coarse dust filter should be cleaned once a week and changed
no less than every 6 months.
The fine filter should be changed after no more than 1000 operat-
ing hours (filter change symbol appears in the display).
For reasons of hygiene, we recommend replacing the following parts at the intervals
stated:
pressure measuring hose every 6 months and earlier if soiled
complete mask system every 6 to 12 months depending on soiling
exhalation system in accordance with the relevant operating
instructions
In addition, servicing should be carried out at the following intervals as a preventive
measure:
after every 5000 operating hours (maintenance symbol
appears in the display)
after no more than 2 years (see servicing sticker on rear of unit)
Servicing includes:
filter change
cleaning unit
replacing any defective parts
a complete check of unit functions and pressure indicators

Related product manuals