40 Maintenance
EN
9.2 Filter change
Change coarse dust filter
Use only original filters made by WEINMANN. If third-party filters are used, this will
invalidate the warranty and may cause restrictions in terms of function and biocom-
patibility.
1. If the VENTIclick humidifier is connected, first disconnect it from the unit. This will
prevent water running into the unit when the filter is changed. Please also ob-
serve the operating instructions enclosed.
2. Press on the latch of the filter compartment cover
and lift it off.
3. Take the coarse dust filter out of the filter compart-
ment cover.
4. Insert the clean coarse dust filter in the filter com-
partment cover.
5. Insert the filter compartment cover into the opening
in the housing bottom edge first.
6. Push the filter compartment cover into the housing
until the latch engages.
Change fine filter
The fine filter needs changing when it has turned dark
in colour, but in any event, after no more than 1000
operating hours. In the latter case, the message
"Change filter!" appears in the display if the
VENTIlogic is switched on.
Press the alarm acknowledgement button . The filter
change symbol then appears continuously in the
status line.
Filter change indicator

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals