28 Hygienic preparation
EN
Creased hose WM 24120 (grey) can be steam-sterilized using
equipment to EN 285. Temperature: 134 ˚C, minimum treatment
time 3 minutes. Follow EN 554/ ISO 11134 with regard to vali-
dation and monitoring.
Accessories
To disinfect the exhalation system, the mask, the VENTIpower, the VENTI-O
2
, the bac-
teria filter and the VENTIclick, please see the section entitled "Disinfection and steriliza-
tion" in the relevant operating instructions.
6.4 Change in patients
If the unit is being operated with a bacteria filter, please note the following:
change the bacteria filter
or
sterilize the bacteria filter and change the particle filter located in-
side it.
If the unit is to be utilized for another patient without using a bacteria filter, it must
undergo hygienic preparation beforehand. This has to be carried out by the manu-
facturer (WEINMANN) or a dealer.
The procedure used for hygienic preparation is described in the inspection sheet and
in the service and repair instructions for the VENTIlogic.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic BiLevel ST and is the answer not in the manual?

Weinmann VENTIlogic BiLevel ST Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic BiLevel ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals