26 Hygienic preparation
EN
The remaining drying time is displayed. The VENTIlogic switches off automati-
cally after 30 minutes. You can interrupt the process at any time by keeping the
menu switch , the dial or the on / off switch depressed for 2 seconds. If
the hose system is still damp in places after drying, start the drying process
again.
4. Remove the drying adapter from the unit outlet.
Clean the VENTIlogic housing
Warning!
There is a risk of electric shock. Before cleaning, remove the con-
necting cord from the supply connection on the rear of the unit and
the supply plug from the supply socket without fail.
Make sure no liquids get into the unit. Never immerse the unit in dis-
infectant or other liquids. Otherwise the unit may be damaged, re-
sulting in a hazard to users and patients.
1. Wipe the unit and the connecting cord clean with a soft damp cloth. The
VENTIlogic must be completely dry before being started up.
2. Remove the filter compartment cover.
3. Remove the coarse dust filter as described in 9.2.
4. Clean the filter compartment cover under running
water so no residues remain. Then dry it carefully.
5. Replace the coarse dust filter and the filter compartment cover as described in
9.2.
Clean coarse dust filter/change fine filter
1. Remove the filter compartment cover as described in 9.2.
2. Remove the coarse dust filter from the filter compartment cover and clean it under
clear running water so no residues remain.
3. Change the fine filter if required.