Use
38 / 96 17.10.2019
5.2.5 Programming a quick test
Test pressure
ü The ProofMaster shows the home screen.
a) Press .
ð The Quick Select screen appears.
b) Tap Pressure.
ð The selection of the test pressure values appears.
c) Press the desired test pressure or press Manual.
ð The test pressure is programmed or a number field appears
when Manual is selected.
d) For an individual test pressure, enter the desired test pressure
value between -0.7 bar and 10.0 bar and then press to
confirm the entry.
ð The individual test pressure is programmed.
e) Press Watch size to program the size of the watch.
ð The selection of the watch sizes appears.
f) Press the value corresponding to the size of the watch.
g) If necessary, press Vacuum test to perform a vacuum test before
the pressure test.
Info
The pressure for the vacuum test is fixed.
The pressure for the vacuum test is -0.3 bar and cannot be changed.
ð The Vacuum test button is displayed in blue.