Use
17.10.2019 57 / 96
Abbreviation Program name Pressure values
(bar)
Limit values
(µg/min)
Watch size
(mm)
CL Calibrated Leak 2.0 50 20–40
P01 Safe test -0.2/2.0 200/200/50 20–40
P02 Standard Men -0.4/3.0 200/50 20–40
P03 Ladies -0.4/2.0 200/50 < 20
P04 Jewelery -0.2/1.0 200/50 < 20
P05 Diver 100 m -0.7/1.0/10.0 200/200/50 20–40
Info
The 50 µg/min limit is only applied at the highest test
pressure of a program, as per ISO 22810. The previous test
pressures are tested for the preliminary test with a limit of
200 µg/min.
5.5.3 Creating a program
Creating a program
a) Press to display the program menu options.
b) Press to open the program menu options.