90 ba75964e04 04/2015
Man. temperature -25 ... +25 ... +130 °C Entry of the manually deter-
mined temperature. For
measurements without tem-
perature sensor only.
Alternative tempera-
ture
on
off
Takes the measured tem-
perature value from an IDS
sensor.
This menu item is only avail-
able if an IDS adapter and
an IDS sensor with temper-
ature sensor are connected.
ISE setup /
AutoRead criterion
low
medium
high
Selection of the AutoRead
criteria (see section 6.1.1
M
EASURING THE ION CON-
CENTRATION, page 47).
ISE setup /
Ion type
Ag, Br, Ca, Cd, Cl,
CN, Cu, F, I, K, Na,
NO3, Pb, S, NH3,
NH4*, CO2, ION
Selection of the ion type to
be measured.
An ion that is not included in
the list can be measured
with the setting, ION.
* Measuring with the NH 500 electrode:
The NH4 setting is not suitable for the gas-sensitive
electrode NH 500.
Select the following settings:
Ion type "ION", Valency "-1".
ISE setup /
Unit
mg/l
µmol/l
mg/kg
ppm
%
Selection, with which unit
the measurement result and
calibration standards should
be displayed.
ISE setup/
Valency
-8 ... +8 Set the valence (Valency)
and molar weight (Molar
mass) for the ion (only with
Unit: mg/kg, ppm, %)
ISE setup/
Molar mass
1 ... 300 g/mol
ISE setup/
Density
0.001 ... 9.999 g/ml or
kg/l
Adjustable density of the
test sample (only with Unit:
mg/kg, ppm, %)
Method Standard addition
Standard subtraction
Sample addition
Sample subtraction
Blank value addition
Selection of the available
measuring methods.
Start method Start measurement with the
selected method.