<1. General>
1-3
IM 12B07W01-04EN 1st Edition : Nov. 11, 2016-00
Note 2: The wiring and piping between the chemical cleaning unit, sensor, KCl reserve tank, and holder are to
be done by the user.
Note 3: The wiring between the switch in the holder and the chemical cleaning unit is to be done by the user.
The terminal size is M4.
EMC Compliance: Korea Electromagnetic Conformity Standard Class A
한국 전자파적합성 기준
A급 기기 (업무용 방송통신기자재)
 이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
n PH8SM3 Holder
Functions:
Moves the pH sensor up or down using a pneumatic cylinder (air cylinder), and provides
chemical cleaning.
Structure: Constructed for indoor use (
For outdoor installation, select Cylinder cover /SC as Option.)
Max. External Dimensions:
190 (W) x 170 (D) x Approx. 830 to 2030 mm (H) (when the sensor is raised)
Mounting:
Mounted on a 2-inch (50A) vertical pipe (outside diameter: 60.5 mm). Two mounting
brackets are supplied.
pH Sensor Up/Down Movement (nominal):
300 mm, 600 mm, 1000 mm, 1500 mm
Weight:
Approx. 8 kg (holder with 300 mm movement)
Approx. 10 kg (holder with 600 mm movement)
Approx. 12 kg (holder with 1000 mm movement)
Approx. 15 kg (holder with 1500 mm movement)
Materials
Frame: Baking nish over staineless steel
Color: Deep sea moss green (Munsell 0.6GY3.1/2.0 or equivalent)
Bottom Cover Pull-up Mechanism: Rod (PPS resin) and screws (PEEK)
Mounting bracket: staineless steel
Holder: Polypropylene and hard PVC (for part of a holder)
Solution Chamber: Hard PVC
O-ring: Fluororubber
Gasket: EPDM
Solution Temperature Range: -5 to 80°C
Ambient Temperature: 0 to 45°C
Flow Speed: 2 m/s or less
Provide an adequate ventilation when using chemicals.