<2. Notes on Handling>
2-8
IM 01C50G01-01EN
2.7.2 IECExCertication
(1) Technical Data
a) IECEx intrinsically safe approval
Caution for IECEx intrinsically safe approval.
Note 1. Certication information
4 - 20mA type
YTA610 and YTA710 with /SU2 temperature
transmitter (4 - 20mA type) is applicable for use
in hazardous locations.
Applicable Standard:
IEC 60079-0: 2011, IEC 60079-11: 2011
Certicate No. IECEx FMG 16.0014X
Type of protection and marking code:
Ex ia IIC T5…T4 Ga
Ex ic IIC T5…T4 Gc
Ambient Temperature:
–40 to 70°C for T4, –40 to 50°C for T5 (Ex ia)
–30 to 70°C for T4, –30 to 50°C for T5 (Ex ic)
Enclosure: IP66/IP67
Overvoltage category: I
Electrial parameters (Ex ia):
Supply/Output circuit: Terminals: +, -
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF,
Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF,
Lo=3.9mH
Electrial parameters (Ex ic):
Supply/Output circuit: Terminals: +, -
Ui=30V, Ci=22nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF,
Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min
(See specic conditions of use)
Fieldbus type
YTA610 and YTA710 with /SU25 temperature
transmitter (Fieldbus type) is applicable for use
in hazardous locations.
Applicable Standard:
IEC 60079-0: 2011, IEC 60079-11: 2011
Certicate No. IECEx FMG 16.0014X
Type of protection and marking code:
Ex ia IIC T4 Ga
Ex ic IIC T4 Gc
Ambient Temperature(Ex ia): –55 to 60°C
Ambient Temperature(Ex ic): –30 to 60°C
Enclosure: IP66/IP67
Overvoltage category: I
Electrical parameters (Ex ia):
Supply/Output circuit: Terminals: +, -
FISCO eld device or
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF,
Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF,
Lo=3.9mH
Electrical parameters (Ex ic):
Supply/Output circuit: Terminals: +, -
FISCO eld device or
Ui=30V, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF,
Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min
(See specic conditions of use)
WARNING
Specic conditions of use
Electrostatic charges on the non-metallic
parts (excluding glass parts) or coated parts
of the Temperature Transmitter shall be
avoided.
When the enclosure of the Temperature
Transmitter is made of aluminum alloy, if it is
mounted in an area where the use of EPL Ga
equipment is required, it must be installed
such that, even in the event of rare incidents,
an ignition source due to impact and/or
friction sparks is excluded.
The dielectric strength of 500 V r.m.s.
between the intrinsically safe circuit and the
enclosure of the Temperature Transmitter
is limited, only by the removable surge
absorber F9220AR.
Note 2. Note for multiple types of protection
(SU2 and SU25)
For the installation of this transmitter, once a
particular type of protection is selected, any
other type of protection cannot be used. The
installation must be in accordance with the
description about the type of protection in this
instruction manual. Cross out the unnecessary
type of protection on the name plate in the
same ways of ATEX.
Note 3. Installation
Installation should be in accordance with
Control Drawing IIE029-A63.