<7. General Specications>
7-4
IM 01C50G01-01EN
Communication Requirements (BRAIN Type)
Communication Distance
Up to 2 km (1.25 miles) when using CEV
polyethylene-insulated PVC-sheathed
cables. Communication distance varies
depending on type of cable used.
Load Capacitance
0.22 μF or less
Load Inductance
3.3 mH or less
Input Impedance of communicating device
10kΩ or more at 2.4 kHz
Load Requirements (HART and BRAIN Type)
0 to 1290Ω for operation
250 to 600Ω for digital communication
PhysicalSpecications
Enclosure
Material & Coating
• Low copper cast aluminum alloy
[for aluminum housing]
Urethane-cured polyester powder coating
Mint-green paint (Munsell 5.6BG 3.3/2.9
or its equivalent)
[for option code /P or /X2]
Epoxy and polyurethane resin solvent
coating
• ASTM CF-8M Stainless steel
Degrees of Protection
IP66/IP67, TYPE 4X
Name plate and tag
316 SST
Mounting
Optional mounting brackets can be used
either for two-inch pipe or at panel
mounting.
Terminal Screws
M4 screws
Integral Indicator (with indicator model)
5-digit numerical display, 6-digit unit display and
bar graph.
Local Parameter Setting (with indicator model)
Parameter conguration by the push button
offers easy and quick setup for parameters.
Accessible parameters are different with each
output cord.
Weight
Alminum housing:
1.3 kg (2.9 lb) without integral indicator and
mounting
Integral indicator: 0.2 kg (0.4 lb)
Bracket for horizontal pipe: 0.3 kg
Bracket for vertical pipe: 1.0 kg
Stainless housing:
3.1 kg (6.8 lb) without integral indicator and
mounting
Integral indicator: 0.3 kg (0.7 lb)
Connections
Refer to “Model and Sufx Codes.”