EasyManua.ls Logo

Zeiss Axiostar plus - Attaching Condensers; Attaching Abbe Condenser 0.9;1.25; Inserting Slider for Phase Contrast or Darkfield

Zeiss Axiostar plus
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Start-Up
Carl Zeiss Attaching condensers Axiostar plus
2-14 B 40-815 e 12/01
1 Condenser carrier
2 Spring pin
3 Centering screws
4 Groove
5 Orientation screw
6 Aperture diaphragm lever
7 Condenser
Fig. 2-12 Attaching Abbe condenser 0.9/1.25
1 Condenser
2 Guiding slot
3 Slider for phase contrast or darkfield
4 Orientation screw
Fig. 2-13 Inserting slider for phase contrast
or darkfield
2.8 Attaching condensers
2.8.1 Attaching Abbe condenser
0.9/1.25
• Unscrew both centering screws (2-12/3) on
the condenser carrier (2-12/1) until the
Abbe condenser 0.9/1.25 (2-12/7) can be
easily inserted into the condenser carrier.
• Insert the condenser into the condenser
carrier in such a way that
1. the dovetail of the condenser is pressed
against the spring pin (2-12/2) in the
condenser carrier and
2. the orientation screw (2-12/5) on the
underside of the condenser enters the
groove (2-12/4) on the condenser
carrier.
• Tighten both centering screws (2-12/3) on
the condenser carrier (2-12/1) until they
engage in the dovetail and keep the
condenser (2-12/7) in position.
2.8.2 Inserting slider for phase
contrast or darkfield
• Insert the slider (2-13/3) until it engages in
the opening in the condenser (2-13/1); the
orientation screw (2-13/4) must face
upwards and engage in the guiding slot
(2-13/2).