Start-Up
Axiostar plus Attaching condensers Carl Zeiss
B 40-815 e 12/01 2-15
2.8.3 Attaching Abbe condenser with
5-position turret
Attachment is made in the same way as the
Abbe condenser 0.9/1.25.
• Loosen both centering screws (2-14/3) on
the condenser carrier (2-14/1).
• Press condenser with dovetail against the
spring pin (2-14/2) in the condenser carrier.
Make sure that the orientation screw
(2-14/5) on the condenser engages in the
groove (2-14/4) of the condenser carrier.
• Tighten both centering screws (2-14/3) on
the condenser carrier (2-14/1) until they
engage in the dovetail and hold the
condenser (2-14/7) in position.
1 Condenser carrier
2 Spring pin
3 Centering screws
4 Groove
5 Orientation screw
6 Aperture diaphragm lever
7 Condenser
Fig. 2-14 Attaching Abbe condenser with 5-position
turret