ECG signal measurement is intended for users of different ages, but for the children, if
there is crying and other uncoordinated factors, it will cause interference and ECG artifacts,
which should be based on the diagnosis of the clinician.
When use the unit on a pacemaker patient, please activate the PACE Detection with
reference to Section 4.2.2.
Wherever the ECG machine is used, you must verify that it is connected with
reliable dedicated grounding wire.
1.3.11 Safety Standards and Requirements
Strictly in accordance with EN 60601-1:2006+A1:2013 Medical electrical equipment - Part
1: General requirements for basic safety and essential performance; IEC60601-2-25
Medical electrical equipment-Part1-2:General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility-
Requirements and tests. The security type is type I CF.
Power supply conditions for the room where the ECG machine is to work shall be suitable
for standard three-plug socket with the grounding plug properly grounded. Otherwise, you
are required to ground the unit using the accompanying grounding cable, with one end
connected to the grounding post and the other end to the ground.
Grounding shall be performed in accordance with relevant standards or under
the guidance of experienced electricians.
When the unit is used in combination with other medical devices, you are required to
connect the grounding cable together with that of the others so as to protect the patient
from possible shock.
Connect one end of the grounding cable to the equipotential post of the ECG machine and
other end to the ground. Avoid using water piping or other piping as a grounding conductor,