31
If the device is damaged or does not work, it cannot be used to acquire and
record the patient ECG. Please contact the service personnel or the
manufacturer immediately.
3.5.2 Power on
After installation and checks, connect the power cable and turn on the ECG machine, then start
to acquire and record the patient's ECG.
3.6 ECG Machine Setting
There are some settings need to be done for the first time use, such as time, report storage
modes etc. During ECG waveform acquisition and measurement, you need to set such
parameters as recording modes, printing duration, gain, paper speed, filter etc. For more details,
please refer to Section 2.3 and Chapter 4.
3.7 Power Off
Please follow the steps below to power off the ECG machine:
1)Verify that the patient ECG acquisition and recording can be ended.
2)Disconnect the ECG electrodes with the patient.
3)Press the power switch to power off the ECG machine.
If you cannot shut down the machine normally, or special circumstances arise,
please long press the power switch for 10 seconds to force a shutdown.
Forced shutdown may incur ECG data loss, thus not recommended usually
except in special circumstances.