EasyManuals Logo

0N3 X5 System User Manual

0N3 X5 System
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
MANUGENX5 - Version 1.01 24.03.2021
© 2021 0N3 s.r.l. - Subject to change without notice
28
5.3 Handheld device chemical resistance
Test 1 (Less strict)
The units under test (UUT) are placed in a closable plastic box (120 x 85 x 65 mm).
A ball of absorbent cotton appositely tinctured with solvent will be placed above the UUT; to avoid early evaporation,
a generic solid body will be put over the ball or, in a more simply way, the closable plastic box will be closed. After a 10
minutes wait, the eventually body and the ball of absorbent cotton will be removed; the solvent that remains on the
UUT will not be wiped off and the box will be closed immediately afterwards for 24 hours.
The test will be performed at environmental temperature (about 20°C).
Test 2 (Very strict)
The units under test (UUT) are fully and thoroughly wet by solvent, then will be closed into a closable box (120 x 85 x 65
mm) for 24 h. Approximately 5 ml solvent will be sprayed over the UUT. The box will be closed and the UUT will remain
in the closed box for at least 24 hours.
The test will be performed at environmental temperature (about 20°C).
Touchscreen test procedure
The Touchscreen is placed into a closable plastic box (120 x 85 x 65 mm) and a ball of a Base Station or bent cotton
appositely tinctured with solvent will be placed above it, then the box will be closed for 1 h.
The test will be performed at environmental temperature (about 20°C).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 0N3 X5 System and is the answer not in the manual?

0N3 X5 System Specifications

General IconGeneral
Brand0N3
ModelX5 System
CategorySecurity System
LanguageEnglish