EasyManuals Logo

3M GMI PS200 Series User Manual

3M GMI PS200 Series
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
64171, Rev. 10/May 2019
3M™ GMI™ PS200 PORTABLE GAS MONITOR
ALARMS
4-2
Figure 4-2: LEL HIHI Alarm
4.2. Acknowledge Gas Alarms
After the gas readings have returned to a safe level, press and hold the R button to clear all alarms.
4.3. Mute Alarms
If configured, muting the alarm will silence the audible alarm for 60 seconds, as follows:
Non-latching alarm: silences the audible alarm for 60 seconds. If the gas reading during that time
falls below the alarm set point, the visual alarm clears automatically.
Latching alarm: silences the audible alarm for a period of 60 seconds. If the gas reading during
that time falls below the alarm set point, the user must acknowledge the visual alarm to clear it.
4.4. High Flammable Gas Over-range Alarm
An over-range alarm protects the user and the LEL sensor during exposures to high concentrations
of flammable gas. If the LEL reading exceeds 100% LEL (see
Figure 4-3: Over Range Alarm):
monitor vibrates;
displayed value changes to 4 rising arrows;
display turns red;
audible alarm tone changes;
LED’s flash quickly;
DANGER OVERRANGE flashes on the display.
WARNING: HIGH OFF-SCALE READINGS MAY INDICATE AN EXPLOSIVE
CONCENTRATION.

Table of Contents

Other manuals for 3M GMI PS200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3M GMI PS200 Series and is the answer not in the manual?

3M GMI PS200 Series Specifications

General IconGeneral
Brand3M
ModelGMI PS200 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals