3M™ GMI™ PS200 PORTABLE GAS MONITOR
MANUAL BUMP TEST OPTIONS
6-3
6.3. Applying Test Gas
Apply the test gas to the monitor via the direct flow regulator set to 0.5 l/min, as shown in Figure 6-4:
Bump Test Kit.
Figure 6-4: Bump Test Kit
6.4. Quick/Full Bump Test
Operation now depends on whether QUICK or FULL bump is configured.
6.4.1. QUICK BUMP TEST
As the alarm threshold for each range is exceeded, the audible/visual/vibration alarms will activate
and a symbol will appear, otherwise a symbol will be displayed.
6.4.2. FULL BUMP TEST
After a short period of time, the gas readings are checked against configurable limits. The audible/
visual/vibration alarms will activate and a symbol will appear if the readings are within these lim-
its, otherwise a symbol will be displayed.