EasyManua.ls Logo

3M GMI PS200 Series - LEL ZERO Fail

3M GMI PS200 Series
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-2
3M™ GMI™ PS200 PORTABLE GAS MONITOR
WARNINGS & FAULTS
64171, Rev. 10/May 2019
Figure 5-3: LEL ZERO Fail
To acknowledge this fault, press the R button once. This will clear the audible/visual alarms. The
wrench symbol will remain on and ZERO FAULT is displayed as shown in
Figure 5-4: LEL ZERO
FAULT
.
Figure 5-4: LEL ZERO FAULT
Return the monitor to clean air and switch it OFF then ON. If the ZERO FAULT persists, return the
monitor to a 3M GMI approved service/repair facility.
A configurable option only allows the user to switch the monitor OFF if a ZERO FAULT is detected,
as shown in
Figure 5-5: ZERO FAULT – Switch OFF.
Figure 5-5: ZERO FAULT – Switch OFF
5.3.2. ZERO FAULT – DURING OPERATION
If ZERO FAULT is displayed during instrument operation, then:
NOTE: The monitor can still detect and alarm on all other sensors.

Table of Contents

Other manuals for 3M GMI PS200 Series

Related product manuals