EasyManua.ls Logo

A&D FX-2000i - Page 43

A&D FX-2000i
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
Class
Item and Parameter
Description
Cp Hi
Upper limit
Refer to 10.10. Comparator
Function”.
Cp lo
Lower limit
dout
Data output
prt
Data output mode
0 Key mode
Accepts the PRINT key only
when the display is stable.
1
Auto print mode A
(Reference = zero)
Outputs data when the display
is stable and conditions of
ap-p, ap-b and the reference
value are met.
2
Auto print mode B
(Reference = last stable value)
3
Stream mode
Outputs data continuously.
4 Key mode B (Immediately)
Accepts the PRINT key regardless of
the display condition.
5 Key mode C (When stable)
Accepts the PRINT key immediately
when the display is stable, or waits for
the display to be stable when not.
6
Interval output mode
Uses interval output mode.
ap-p
Auto print polarity
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
2
Both
Regardless of displayed value
ap-b
Auto print difference
0
10 digits
Difference between reference
value and displayed value
1
100 digits
2
1000 digits
int
Interval time
0
Every measurement
Interval time for the interval
output mode
(With prt 6)
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minute
7
5 minute
8
10 minute
5-id
ID number output
0
No output
Selects whether or not the ID
number is output.
1
Output
5-td
*
Time/Date output
0
No output
1
Time only
2
Date only
3
Time and date
pU5e
Data output pause
0
No pause
Selects the data output
interval.
1
Pause (1.6 seconds)
at-f
Auto feed
0
Not used
Selects whether or not
automatic feed is performed.
1
Used
info
GLP output
0
No output
Selects GLP output method.
1
AD-8121 format
2
General data format (output time and date using
the built-in clock)
3
General data format (output time and date using
the clock of the external equipment)
ar-d
Zero after output
0
Not used
Adjusts zero automatically
after data is output
1
Used
Factory settings. Digit is a unit of minimum weighing value
*
:
Only for the FZ-i
/
FZ-iWP
/
FZ-GD series

Table of Contents