SPECIFICATIONS
FZ series
120
i
/
i
WP/GD
200
i
/
i
WP/GD
300
i
/
i
WP/GD
500
i
1200
i
/
i
WP/GD
2000
i
/
i
WP/GD
3000
i
/
i
WP/GD
5000
i
Weighing capacity [ g ]
122
220
320
520
1220
2200
3200
5200
Maximum display [ g ]
122.084
220.084
320.084
520.084
1220.84
2200.84
3200.84
5200.84
Minimum weighing value
[ g / digit]
0.001 0.01
Repeatability [ g ]
(Standard deviation)
0.001 0.01
Linearity [ g ]
±
0.002
±
0.02
Stabilization time
(typical at FAST)
Approx. 1 second
Sensitivity drift
(10°C to 30°C/50
°
F to 86
°
F )
±2 ppm/°C
Accuracy right after calibration
using the internal mass.
See notes below *
±
0.010 g
±
0.10 g
±
0.15 g
The value above is to the weighing capacity.
Internal mass
Yes
Clock and calendar function
Yes
Operating environment
5 °C to 40 °C (41 °F to 104 °F)
85 %RH or less (No condensation)
Display refresh rate
5 times/second, 10 times/second or 20 times/second
Counting
mode
Minimum unit mass
0.001 g
0.01 g
Number of samples
5, 10, 25, 50 or 100 pieces
Percent
mode
Minimum 100%
reference mass
0.100 g 1.00 g
Minimum 100%
display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
100 g
50 g
200 g
100 g
50 g
300 g
200 g
100 g
50 g
500 g
400 g
300 g
200 g
100 g
50 g
1000 g
500 g
2000 g
1000 g
500 g
3000 g
2000 g
1000 g
500 g
5000 g
4000 g
3000 g
2000 g
1000 g
500 g
Weighing pan diameter
130 mm
150 mm
Net weight
Approx. 3.0 kg (
iWP series:
approx. 3.2 kg)
External dimensions
193 (W) x 262.5 (D) x 85.5 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface
RS-232C
Dustproof and waterproof specifications
(FZ-iWP / FX-iWP series only)
IP65 standard
* Accuracy right after calibration using the internal mass, with the accessory breeze break
installed and in good ambient conditions (within the temperature range of 10
°
C to 30
°
C
(50
°
F to 86
°
F) with no abrupt changes in temperature or humidity, no drafts, no vibration
and no effect by magnetic fields or static electricity).
With the FZ-i / FZ- i WP / FZ-GD series balance, the environment and aging may cause internal
mass value error. Calibrate using an external mass and perform maintenance regularly.